Advanced MODULAR benchtop meter, allows the simultaneous use of 4 equal or different modules. The possible modules are:
pH/Temperature:
- Up to 5 calibration points with automatic recognition of USA/NIST and DIN standards.
- Correction of calibration errors without recalibration with the edit function.
pH/Ion concentration/Temperature:
- All the benefits of the pH module.
- Advanced functions for ion concentration measurement: selective electrode identification, auto-whitening, isopotential, stability.
- Incremental methods: known addition, known subtraction.
Conductivity/Salinity/Resistivity/TDS/Temperature:
- Selectable reference temperature: 5,10,15, 20 or 25ºC.
- Temperature correction: linear, non-linear, or
Dissolved oxygen/RDO/Temperature:
- Automatic temperature compensation.
- Salinity compensation.
- Barometric pressure compensation (automatic or manual).
The calibration is stored in the module itself.
Color display with:
- Real-time graphics - Alarms and indicators on screen.
- User selectable display settings.
- Numeric keypad. - Easy communication with computer: RS232 output, two USB ports. - Simultaneous display of all modules connected to the equipment. - Non-volatile memory, stores up to 2000 readings referenced with date and time. - Includes analysis software to transfer stored results or readings on screen in real time.
Versa Star Pro meter dimensions: 28cm x 24cm x 10cm (length x width x height)
Weight of the meter: 1.2 kg
Dimensions of the modules for Versa Star: 12 x 3 x 9.5 cm (length x width x height)
Weight of modules: 0.2 kg

Module parameter specifications:
pH MODULE:

  • pH

Range: -2,000 to 20,000
Resolution: 0.1/0.01/0.001
Accuracy: ± 0.002
Calibration points: Up to 6
Calibration editing: Yes

  • mV/RmV

Range: ± 2,000.0 mV
Resolution: 0.1
Accuracy: ±0.2 mV or ±0.05 % of reading
ORP mode: Yes

Temperature
Range: -5 to 105 ºC
Resolution: 0.1
Accuracy: ±0.1
Offset calibration: 1 point

POLAROGRAPHIC OXYGEN DISSOLUTE/RDO/DO MODULE:
O.D.
POLAROGRAPHIC O.D. CONCENTRATION
Range: 0 to 90 mg/l
Resolution: 0.01/0.1
Accuracy: ± 0.2

% SATURATION-POLAROGRAPHIC
Range: 0 to 600
Resolution: 0.1/1
Accuracy: ±2%.

CONCENTRATION OF O.D. RDO LUMINISCENT
Range: 0 to 50 mg/l
Resolution: 0.01/0.1
Accuracy: ± 0.1 up to 8 mg/l // ±0.2 from 8 to 20 mg/l // 10 % of reading from 20 to 50 mg/l

% SATURATION- RDO LUMINISCENT
Range: 0 to 500
Resolution: 0.1/1
Accuracy: ±2 %
Pressure Correction Bar: 450 to 850 mm Hg
Salinity correction factor: Automatic if Conductivity/Salinity module is present. Or manual: 0 to 45 ppm

  • Temperature

Range:0 to 50 ºC
Resolution:0.1
Accuracy: ±0.1
Offset calibration: 1 point

pH/Ion Concentration MODULE:

  • pH

Range: -2,000 to 20,000
Resolution: 0.1/0.01/0.001
Accuracy: ± 0.002
Calibration points: Up to 6
Calibration editing: Yes
Adjustable Isopoint: Yes

  • mV/RmV

Range: ± 2,000.0 mV
Resolution: 0.1
Accuracy: ±0.2 mV or ±0.05 % of reading
ORP mode: Yes

  • Concentration

Range: 0 to 19,999
Resolution: Up to 4 significant digits
Accuracy: ±0.2 mV or ±0.05 % of reading
Units: ppm, M, mg/l, %, ppb.
Calibration points: Up to 6
Calibration editing: Yes
Adjustable Isopoint: Yes
Calibration options: Linear, point-to-point, non-linear with auto-whitening, stability at low concentration.

  • Temperature

Range: -5 to 105 ºC
Resolution: 0.1
Accuracy: ±0.1
Offset calibration: 1 point

CONDUCTIVITY MODULE:
Conductivity
Range: 0.001 to 3000 mS
Resolution: 0.001 µS minimum, 4 digits significant.
Accuracy: 0.5 % of reading
Reference Temperature: 5,10,15, 20 or 25 ºC.
Temperature Compensation: Linear, non-linear,
Calibration: up to 6 points.
Calibration editing: Yes

  • Resistivity

Range: 2 ohm to 100 meg-ohm
Resolution: 2 ohm-cm
Accuracy: 0.5 % of reading

  • Salinity

Range: 0 to 80 ppt NaCl eq
Resolution: 0,01
Accuracy: ±0,1

  • TDS

Range: 0 to 200 ppt
Resolution: 4 significant digits
Accuracy: ±0.5 %
TDS: Linear 0.01 to 10.00

  • Temperature

Range: -5 to 105 ºC
Resolution: 0.1
Accuracy: ±0.1
Offset calibration: 1 point

pH/Temperature MODULE with LogR:

  • pH

Range: -2,000 to 20,000
Resolution: 0.1 – 0.01 – 0.001
Accuracy: ± 0.002
Calibration points: Up to 6
Calibration editing: Yes

  • mV/RmV

Range: ± 2,000.0 mV
Resolution: 0.1
Accuracy: ±0.2 mV or ±0.05 % of reading
ORP mode: Yes

  • Temperature with LogR

Range: 0 to 100 ºC
Resolution: 0.1
Accuracy: ±0.5
Resistance: 0 to 6000 MΩ
Offset calibration: 1 point

  • Temperature without LogR

Range: -5 to 105 ºC
Resolution: 0.1
Accuracy: ±0.1
Offset calibration: 1 point

pH/ISE /Conductivity/RDO/OD/ORP/T

GLP/GMP
CFR 21 part 11 (together with Navigator Pro Software)
IP 54
European Pharmacopoeia Conductivity 2.2.38
European Pharmacopoeia pH 2.2.3

pH electrodes
Calibration standards
Temperature probe
pH buffers
DO probe
RDO probe
ISE electrodes (Ammonium, Sodium, Fluorides, Nitrates, Chlorides, Chlorides, Iodide, Lead, Bromide, Silver, Sulfide, Cadmium …)
Software
Other accessories available on request

Pide información de VERSA STAR Pro series benchtop multiparameter meter