Total Chlorine Analyzer model CLORA 2XP R-Series, standard sample cup

Description

Specifications

Parameters

Standards

Accessories

Descargas

The new analyzer model Clora 2XP R analyzes total chlorine in liquid hydrocarbons such as aromatics, distillates, heavy fuels and crude oil, as well as aqueous solutions, while automatically correcting for sulfur interference according to ASTM D7536 and D4929C.
Test Method: MWD XRF® - High Definition Monochromatic Wavelength X-ray Fluorescence. Applications:
  • Total Chlorine Analysis from aqueous solutions and aromatic products to heavy fuels and crude oils
  • For refineries, petrochemical and additive plants, pipeline terminals and test laboratories
Characteristics:
  • LOD: 0.07 ppm
  • Range:
    • Low: from 0.1 to 3000 ppm
    • High: 0.3 to 2 wt%.
  • Low maintenance: no gases, no heating elements, no columns, no quartz tubes.
  • Automatic sulfur correction(value not reflected on screen only correction).
  • 43 mm traditional XRF sample cups
  • Easy to use:
    Intuitive 10-inch touch screen
    Simply connect and measure
    Measuring time: 10-999 s
  • Dimensions: (Height x Width x Depth): 42 x 40 x 54 cm
  • LIMS integration for data management and transfer
  • Customized sample presets to save data entry time and minimize errors
  • Supports up to 30 calibration curves
  • USB connectivity
  • Advanced error reporting and diagnostics
  • Minimum volume of sample cup: 5 ml

Total Chlorine in liquid hydrocarbons

 

ASTM D7536
ASTM D4929C

To be consulted

Pide información de Total Chlorine Analyzer model CLORA 2XP R-Series, standard sample cup

We want to play an active role in creating a safer and more environmentally respectful future.