Fuel Sulfur Analyzer by fluorescence RX WDXRF wavelength dispersive monochromatic wavelength, model SINDIE 2622 Gen 3.
According to standards:
ASTM D 2622, ASTM D 7039 and ISO 20884
User selects the method. Can work with or without helium for the D2622, the D7039 does not require it.
Test method:
MWD XRF - Monochromatic Optics
Associated for excitation focusing which improves the signal-to-noise ratio.
Focusing of emitted radiation that increases detection limits reduction of matrix effects.
Easy to use:
Intuitive touch screen
As simple as connect and measure
Measuring time: 30-900 s.
Extremely low maintenance: No conversion gases, heating elements, tubes or quartz columns.
Total sulfur analysis: From ultra-low sulfur diesel and gasoline to heavy fuel oil and crude oil.
For use in multiple application areas such as:
Refinery laboratories
Pipeline terminals
Additive plants
Inspection laboratories
Hassle-free sample preparation using Accu-cell Cuvettes
Low and high range calibrations available:
- Low range: 0.15 ppm - 3000 ppm
- High range: 0.3% wt.% - 10% wt.% High range: 0.3% wt.% - 10% wt.
Dimensions (WxDxH): 37 cm x 50 cm x 34 cm
Weight: 32 kg
Power supply: 200-240 VAC, 47-63 HZ at 6.0 Amps
Measuring range: From 0.15 ppm to 10 %.
Calibrations: Up to 16 calibration curves available in memory. Two calibrations cover both gasoline and diesel matrix over the entire dynamic range:
One from 0.15 ppm up to 3000 ppm/w
Other from 0.3 to 10 % /w
Ambient temperature requirements: 5 – 40 °C
Sample volume: 10 ml
Sample cuvette type: Accu-cell
Measuring time: 30 to 900 seconds, user programmable.
The extended time is due to the need, per the standard, for the measurement of the background plus the measurement of the sample.

Polyamide window

Touch screen with selection of:
– Test method
– Selected calibration

– Rank

Communications: Printer, Ethernet: LIMS connectivity (optional)

Typical repeatability (r) and reproducibility (R) values in diesel fuel, at 95%. Measuring time of 300s.
Sulfur concentration (ppm) Repeatability (r) Reproducibility (R)
2 0.3 0.7
5 0.5 0.8
8 0.6 1.0
15 0.8 1.4
100 2 4
500 5 10

Sulfur

ASTM D2622
ASTM D7039
ISO 20884

Autosampler
Sample cuvette holder (8 positions)
LIMS connectivity
Other accessories on request