For clinical laboratory water
Type I water / ultra pure water (UPW)
Ultrapure water polisher
The Select Neptune Ultimate provides a guaranteed high ultrapure water flow rate of 18.2 MΩcm for analytical and life science laboratory applications.
Characteristics:
- Reliable and space-saving desktop or wall-mounted systems
- Dosing rate up to 2 l/min
- Optional external storage tanks of up to 100 liters
- Installation kit and all consumables included for first year's operation
- Color LCD touch screen panel
- Visual and audible alarms included
- USB port for downloading event data and uploading software updates
- Semi-automatic cleaning cycle
- TOC Monitoring
- Point of use 0.2 μm filtration
Additional Select Neptune Ultimate features:
- Ion exchange cartridges incorporating semiconductor grade monospherical mixed-bed deionizing resin with a low TOC leaching profile and a high activity organic adsorption medium
- Submicron filter and ultrafilter to remove particulates, bacteria and endotoxins
- 185nm or UV photo-oxidant for splitting organic compounds into smaller charged ionic species that can be removed by ion exchange
- 254 nm UV to reduce bacteria by more than 99 %.
- Wifi connection
ECO option:
The ECO Option helps to recover up to 50% of the water and save energy.
UNIT:
Dimensions (W x D x H): 310mm x 560mm x 750mm
Maximum shipping weight: 21Kg
Maximum working weight: 29Kg
Installation requirements:
Power supply: Single phase, 110-230 V, +/- 10 %, 50/60 Hz
Feed water: < 20 μS/cm
Maximum TDS: < 14ppm
Minimum inlet pressure: 5psi (0.34bar)
Maximum inlet pressure: 20psi (1.38bar)
Feed water temperature: 1-35°C.
SYSTEM:
LCD display panel: color touch screen
Internal filtration: Ultrafiltration
Use point: 0.2 μm
Ultraviolet lamp: 185 nm and 254 nm
Recirculation pump: ✓
Dispensing modes: Latched – hold – volumetric
Ultra-pure polishing cartridges: ✓
Dispensing rate: up to 2.0 l/min.
TREATED WATER:
Inorganic: > 18.2 MΩ.CM
pH *: Neutral
Bacteria: < 0.1 cfu/ml
Organics - TOC (ppb): < 1
Particles: Ultrafiltration
Endotoxins: < 0.001 EU/ml
DNAases: < 4 pg/μl
RNases: < 0.01 ng/ml
*The pH of stored water may decrease due to the absorption of free carbon dioxide.
Water Purification
Reverse Osmosis