It has a backlit LCD display that simultaneously shows the readings of any of the parameters with the temperature, as well as the date, time, cell status and calibration points.
Easy operation through guided on-screen menus with key indicators for the user. Software in Spanish.
AUTO-READ function.
Stable reading display.
Up to 5 pH calibration points with recognition of USA / NIST and DIN standards.
Automatic recognition of Polarographic and optical/luminescent RDO probe.
Selectable cell constant for use with 2 or 4 electrode conductivity probes.
Selectable reference temperature 5, 10, 15, 15, 20 or 25ºC, with linear or non-linear correction.
Includes standard conductivity/salinity conversion curve.
Conductivity/TDS conversion curves, linear or IS/EN27888.
Stores up to 5000 readings, associated with date and time, in its non-volatile memory.
Easy download of the data to the PC, through USB port or RS232 port with the analysis software.
It can be battery operated (4 AA alkaline batteries), which provide an autonomy of 800 hours of operation.
It is also possible to purchase the mains adapter (optional) to connect it to 220V.
Portable and submersible with IP67 protection.
Model 3295 Includes:
STAR A 329 Meter
ROSS Triode pH/ATC pH electrode with 3 meter cable (OR8107UWMMD)
DuraProbe Conductivity Probe with 3 meter cable (OR013010MD)
Luminescent oxygen sensor (OR087010MD) with 3 meter cable and stainless steel guard.
pH 4 Buffer, pack 10 sachets (OR910410)
pH 7 Buffer, pack 10 sachets (OR910710)
pH 10 Buffer, pack 10 sachets (OR911010)
Storage solution for pH electrode (OR810001)
Conductivity standard 1413 µS, pack 10 sachets (OR01100710)
Washing solution, pack 10 sachets (OR911110)
Protective frame with electrode holder.
Carrying case
AA batteries (installed)
Dimensions: 24cm x 10cm x 6.5cm (length x width x height)
Weight: 0.5 kg
Power supply: AA batteries (installed)
Autonomy: 800hrs.
Tickets:
Conductivity and temperature probe: 8-pin-mini-DIN
pH / ISE Electrodes: BNC / Pin
DO sensor: 9 pin mini DIN
Outputs: USB, RS232
Measuring modes: pH, mV, relative mV (RmV) or ORP with temperature; conductivity, TDS, salinity or resistivity with temperature; dissolved oxygen as % saturation or mg / L with temperature and barometric pressure
Calibration points: 1-point relative mV (RmV) mode, 1-point ORP mode
Calibration points (conductivity): up to 5
Calibration points (ISE): 1 to 5
Calibration points (pH): 1 to 5
Cell constant: 0.001 to 199.9cm-1
Channels: 3
Operating temperature: 5°C to 45°C (ambient)
Temperature probe calibration: 1-point temperature compensation calibration
Temperature range: -5.0°C to +105.0°C
Temperature resolution: 0.1°C, 0.1°F
Temperature selection: manual or automatic with ATC temperature probe
Parameter specifications:
pH:
Range: -2,000 to 20,000
Resolution: 0.1 / 0.01 / 0.001
Accuracy: ± 0.002
Calibration points: Up to 5 points.
Edit Calibration: Yes.
mV/RmV:
Range: ± 2000.0 mV
Resolution: 0.1
Accuracy: ± 0.2 mV or 0.05% of reading
REDOX mode: Yes.
Temperature:
Range: -5 to 105 ºC
Resolution: 0.1
Accuracy: ± 0.1
Offset compensation: 1 point
ISE:
Range: 0 to 19999
Resolution: Up to 3 significant digits
Accuracy: ± 0.2 mV or 0.05% of the reading
Units: ppm, M, mg/L, %, ppb, none, ppb, none
Calibration points: Up to 5 points.
Edit Calibration: Yes
Calibration data: Programmed end point, linear point-to-point, non-linear with selectable auto white and low range stability.
Conductivity:
Range: 0.001 uS to 3000 mS
Resolution: 0.001 µS minimum 4 significant digits
Accuracy: 0.5% reading ±1 digit
Temperature Ref.: 5,10,15,20,25 ºC
Temperature Compatibility: Linear (0 to 10.0%) Non-linear factor (nLn).
Compatible Cell Constant: 0.001 to 199.9
Calibration Points: Up to 5 points
Edit Calibration: Yes
Resistivity:
Range: 2 ohm to 100 meg-ohm.
Resolution: 2 ohms-cm.
Accuracy: 0.5% reading ±1 digit
Salinity.
Range: 0.01 to 80.0 ppt NaCl, equivalent; 0.01 to 42 ppt, practical salinity
Resolution: 0.01
Relative accuracy: ± 0.1
TDS.
Range: 0 to 200 ppt
Resolution:4 significant digits
Accuracy: ±0.5% reading ±1 digit
Adjustable TDS factor: Linear 0.01 to 10.00
Dissolved Oxygen:
Concentration – Polarographic:
Range: 0 to 90 mg/l
Resolution: 0.01 – 0.1
Accuracy: ±0.2
% Saturation – Polarographic:
Range: 0 to 600
Resolution: 0.1 – 1
Accuracy: ±2%.
Concentration – RDO:
Range: 0 to 50 mg/l
Resolution: 0.01 – 0.1
Accuracy: >8 mg/l: ±0.1
From 8 to 20 mg/l: ±0.2
From 20 to 50 mg/l: 10% of the reading
% Saturation – RDO:
Range: 0 to 500
Resolution: 0.1 – 1
Accuracy: ±2%.
Automatic barometric pressure correction: 450.0 to 850.0 mm Hg
Manual correction of salinity factor: 0 to 45 ppt
Calibration: Air-saturated water, water-saturated air, manual (Winkler) and zero.
Probe: Polarographic or RDO.
Temperature:
Range: -5º to 105ºC, 23 to 221ºF
Resolution: 0.1
Accuracy: ± 0.1
Offset calibration:1 point