Videos
Test methods: 3 integrated test methods, user-programmable via PC Supplied with ASTM D7619; IP 565; ISO 4406 pre-installed Particle size range: ISO 11171: 4 µm (c) a> 100 µm (c) (calibration for larger sizes available upon request)
Test duration (ASTM D7619; IP 565): less than 3 minutes
Pressure (max.): Inline pressure 10 bar, pressure reducing valve for high pressure max. 300 bar
Sample temperature range: 0 to 70 °C
Operating temperature range: 5 to 40 °C
Relative humidity (max.): 80% to 40 ° C
Measuring modes: Individual or continuous, with variable cycles.
Measuring channels: 6 channels are displayed, up to 16 programmable via PC
4 µm (c), 6 µm (c), 10 µm (c), 14 µm (c), 21 µm (c), 25 µm (c), 30 µm (c), 38 µm (c), 70 µm (c) Counts per measurement (max.): 600,000 per ml
Coincidence error limit: 60,000 particles / ml ≥4µm (c) with ≤ 5% coincidence error (ISO 11171). Sample viscosity (max.): 64 mm2 / s (with internal pump), 200 mm2 / s
(pressure fed at 3 BarG) (sample delivery system SA1950-0 is available as an accessory)
Sample volume: 80 ml for ASTM D7619 and IP 565, 20 ml for other methods (including wash cycles)
Sample flow rate: 30 ml / min ± 5 ml / min
Dimensions (HxWxD): 250 x 145 x 350 mm
Weight: 7.4 kg
PC connection: Direct, with supplied software (Optional)
Particulate matter concentration in fuel and oil fluids, hydraulic or used
Certified testing fluid
Sample bottles with stopper
Other accessories on request
Pide información de Particle counter model AvCount Lite
We want to play an active role in creating a safer and more environmentally respectful future.
Related resources
INSTRU joins the United Nations Global Compact
Continue reading
Dew Point measurement: how to control residual moisture in industrial processes
Continue reading
Our thanks for the Marta Mayordomo Award
Continue reading
New reference standards certified at –40 °C for Jet Fuel testing
Continue reading
InnovaPrep solutions for microbiological detection in critical environments
Continue reading
ASTM D93-25: Standard update to improve traceability of the flash point
Continue reading