Particle counter model AvCount Lite

Description

Specifications

Parameters

Standards

Accessories

Equipment for particle concentration determination in fuel and oil liquids, hydraulic or used.
- Fully automatic equipment.
- Direct bottle suction, with programmable cleaning cycles.
- Optionally, a bottle shaking device is available for repetitive results.
- Validation and calibration standard available with certificate.
- Specifically designed according to IP 565-98 Standard, included in DEF-STAN 91 rev.6.

Test methods: 3 integrated test methods, user-programmable via PC                                                                         Supplied with ASTM D7619; IP 565; ISO 4406 pre-installed                                                                  Particle size range:  ISO 11171: 4 µm (c) a> 100 µm (c) (calibration for larger sizes available upon request)   
Test duration (ASTM D7619; IP 565): less than 3 minutes     
Pressure (max.): Inline pressure 10 bar, pressure reducing valve for high pressure max. 300 bar
Sample temperature range: 0 to 70 °C     
Operating temperature range: 5 to 40 °C
Relative humidity (max.): 80% to 40 ° C     
Measuring modes: Individual or continuous, with variable cycles. 
Measuring channels: 6 channels are displayed, up to 16 programmable via PC   
                             4 µm (c), 6 µm (c), 10 µm (c), 14 µm (c), 21 µm (c), 25 µm (c), 30 µm (c), 38 µm (c), 70 µm (c)              Counts per measurement (max.):  600,000 per ml 
Coincidence error limit: 60,000 particles / ml ≥4µm (c) with ≤ 5% coincidence error (ISO 11171).                    Sample viscosity (max.): 64 mm2 / s (with internal pump), 200 mm2 / s   
                                               (pressure fed at 3 BarG) (sample delivery system SA1950-0 is available as an accessory)       
Sample volume: 80 ml for ASTM D7619 and IP 565, 20 ml for other methods (including wash cycles)   
Sample flow rate: 30 ml / min ± 5 ml / min       
Dimensions (HxWxD): 250 x 145 x 350 mm
Weight: 7.4 kg   
PC connection: Direct, with supplied software (Optional)

Particulate matter concentration in fuel and oil fluids, hydraulic or used

ASTM D975; ASTM D7619
IP 565; IP PM FA
DEF STAN 91-091 ; DEF STAN 91-86
ISO 4406

Certified testing fluid

Sample bottles with stopper

Other accessories on request

Pide información de Particle counter model AvCount Lite

We want to play an active role in creating a safer and more environmentally respectful future.