Indicated Cetane Number (ICN) vs. Cetane Index (CI)
Continue readingDescription
Specifications
Parameters
Accessories
Descargas
Mounting options:
Wall or frame mounted, for indoor applications
In enclosure with insulation: Standard fiber or polyester IP66
Dimensions: On frame: 60x50x15cm
In cabinet: 75x50x30cm
User Interface:
Keypad with 4 keys and 4 indicator LEDs
Configurable menus in several languages
Display:
Monochrome backlit – 8 lines x20 characters –
Color, backlit, graphic and panoramic (optional)
Memory:
Microprocessor with internal program (firmware) upgradeable via mini USB or MicroSD.
Logging of 64 analyses, 16 calibrations and 32 alarms or errors.
Communications:
Two 4-20mA analog outputs, separately configurable and galvanically isolated.
A two-wire RS485 bus output.
Optionally the RS485 can use MOD-BUS or PROFIBUS protocol
Two digital inputs assigned to sample missing and reagent missing detection.
Five free assignable digital inputs.
Two free assignable digital outputs.
USB. Loading of programs and I/O data.
uSD card data and I/O program.
Relays:
Four relays with three contacts (C, NO and NC), potential free and programmable.
Calibration:
Manual or Automatic programmed.
Washes (programmable):
Automatic after each test.
With the sample itself.
With rinsing solution
Dosing System:
Syringe dosing system
Fluid storage loop system that allows the syringe to never touch the sample or reagents.
Highly reliable solenoid valves with closing elements in Kalrez.
Large pitch diameter (1.5mm) on solenoid valves.
Environmental Conditions:
Ambient Temperature: -20ºC to 65ºC.
Maximum Relative Humidity: 95% non condensing
Copper (Cu) / Ammonia (NH3) / Nitrogen dioxide (NO2) / Nitrate (NO3) / Phosphate (PO4) / Nickel (Ni) / Chromium (Cr) / Chlorine (Cl2) / Zinc (Zn) / Fluoride (F) / Iron (Fe) / Silica (SiO2) / Sulfur (S2-) / Phenol (C6H6O) / Cyanide (CN-) / Aluminum (Al) / Arsenic (As) / Boron (B) / Chloride (Cl-) / Sodium (Na)
Consult
Pide información de On-line analyzer for physical and chemical parameters in liquid samples
We want to play an active role in creating a safer and more environmentally respectful future.
Related resources
Indicated Cetane Number (ICN) vs. Cetane Index (CI)
Continue reading
Efficient and Non-Corrosive Disinfection Technology for the Pharmaceutical Industry
Continue reading
Hands-off accuracy: the silent revolution in online ammonium analysis
Continue reading
A. Krüss Optronic Refractometers: Optical Precision for Science and Industry
Continue reading
New generation of high-precision online concentration monitors
Continue reading
Oxygen measurement in potentially explosive environments
Continue reading