Dew Point measurement: how to control residual moisture in industrial processes
Continue readingDescription
Specifications
Parameters
Standards
Accessories
Digital refractometer nD ranges 1,3000-1,5800; 0-95 %Brix. Audit Trail
The new DR7000 series are designed to determine refractive index, Brix percentage and other concentrations in liquid samples. Equipped with a CCD optical sensor, Peltier temperature control, electronic measurement and evaluation systems and a touch screen, they have all the essential elements to ensure accurate and reliable measurements. Operation is carried out directly on the device itself, without the need for a separate PC with additional software.
These devices, together with their firmware, comply with all the standards required in the areas regulated by the FDA (depending on the model), as well as by the European (Ph. Eur.) and American (USP) pharmacopoeia. In particular, the DR7200-P and DR7400-P models are designed to comply with the requirements set forth in CFR21 Part 11.
DR7200-P:
| Technical data Refractometer DR7000 series | ||||
| Specification | DR7100-P | DR7200-P | DR7300-P | DR7400-P |
| SCALES | Refractive index (nD) Sucrose concentration, (%Brix) User defined (unlimited) |
|||
| MEASURING RANGE | nD 1.3000 – 1.5800 0-95 %Brix |
nD1,3000 – 1,5800 0-95 %Brix |
nD 1,30000 – 1,72000 0-95 %Brix |
nD 1,30000 – 1,72000 0-95 %Brix |
| MEASURING ACCURACY 1) | nD ±0.0001 ±0.1 %Brix |
nD ±0.00002 ±0.02 %Brix |
||
| RESOLUTION | nD 0.0001 0.1 %Brix |
nD 0.00001 0.01 %Brix |
||
| AUDIT TRAIL FUNCTIONALITY | None | Fully available | None | Fully available |
| MEASUREMENT TIME 3) | Approx. 2 seconds | |||
| MEASURING PRISM | Sapphire | |||
| LIGHT SOURCE | LED 589 nm | |||
| SAMPLE VOLUME/FILL QUANTITY | ≈ 0.3 ml | |||
| SAMPLE WETTED MATERIALS | Sapphire, PTFE and stainless steel 1.4404 (X2CrNiMo 17-12-2) | |||
| Temperature data | ||||
| Specification | DR7100-P | DR7200-P | DR7300-P | DR7400-P |
| TEMPERATURE CONTROL | Integrated Peltier sample temperature control system | |||
| TEMPERATURE RANGE 2) | 10-60 °C | 10-80 °C | ||
| TEMPERATURE ACCURACY | ±0,1 °C | ±0,1 °C | ±0,03 °C | ±0,03 °C |
| TEMPERATURE COMPENSATION | automatic temperature compensation for %Brix according to ICUMSA selectable Temperature compensation can be created on the customer’s side optional |
|||
| TEMPERATURE MEASUREMENT | With integrated Pt100 temperature sensor | |||
| TEMPERATURE MEASUREMENT RANGE | 10-80 °C | |||
| TEMPERATURE MEASUREMENT ACCURACY | ±0,1 °C | |||
| RESOLUTION OF TEMPERATURE MEASUREMENT | 0,1 °C | 0,01 °C | ||
| AMBIENT TEMPERATURE | 10-40 °C | |||
| AMBIENT AIR HUMIDITY | 10-90 % (non-condensing) | |||
| Basic equipment data | ||||
| OPERATION | Stand-alone device (including firmware): 7.0 inch capacitive touch screen, 800 x 480 pixels | |||
| ELECTRONIC COMMUNICATION PORTS | 2x USB (2.0; type A); 1x RS-232; 1x LAN | |||
| HOUSING | Aluminum, powder-coated; protection code: IP20 | |||
| DIMENSIONS (W x H x D) | 230 mm x 170 mm x 360 mm | |||
| WEIGHT OF THE DEVICE | 5 kg | |||
| OPERATING VOLTAGE | 90-240 VAC | |||
| POWER CONSUMPTION | 40 W in measurement mode 1) ; 75 W max. | |||
| DESIGN FREQUENCY | 50-60 Hz | |||
¹) Under normal refractive index measurement conditions (λ = 589 nm, 20 °C, 1013 hPa, 50 % relative humidity)
²) For non-condensing ambient conditions
³) After temperature equalization
Refractive index (nD)
Sucrose concentration, (%Brix)
User defined (unlimited)
Consult
Pide información de New DR7000 Series Digital Refractometer, Model DR7200-P
We want to play an active role in creating a safer and more environmentally respectful future.
Related resources
Dew Point measurement: how to control residual moisture in industrial processes
Continue reading
Our thanks for the Marta Mayordomo Award
Continue reading
New reference standards certified at –40 °C for Jet Fuel testing
Continue reading
InnovaPrep solutions for microbiological detection in critical environments
Continue reading
ASTM D93-25: Standard update to improve traceability of the flash point
Continue reading
High-precision temperature control with a commitment to the environment
Continue reading