Multiparameter benchtop meter model STAR A215 - Complete Kit

Description

Specifications

Parameters

Accessories

Benchtop pH and conductivity meter suitable for all types of samples.
It has a backlit LCD display that shows readings of:
pH / mV / RmV or Conductivity / TDS / Salinity
Simple operation through guided on-screen menus with key indicators for the user.
Software in Spanish.
5 pH calibration points with automatic recognition.
AUTO-READ function.
Stable reading indicator.
Selection of 2 or 4 cell conductivity probe.
Stores up to 2000 readings, associated with date and time, in its non-volatile memory.
Easy download of data to PC via USB port or RS232 port with the analysis software.
Possibility of incorporating a rod stirrer (code 096019, sold separately).
It works with a 220V mains adapter, but can also be powered with batteries (4 alkaline AA type).
IP54 splash-proof protection.
 
The STAR A2155 model includes:
STAR A 215 meter
ROSS pH Triode pH/ATC Electrode (OR8157BNUMD)
Conductivity probe (OR013005MD)
ROSS Solution Kit (pH 4, 7 and 10 buffers, storage solution and cleaning solution, in 475 ml bottles)
Conductivity standard 1413 µS (5 x 60 ml)
Probe holder support arm
Network adapter
Dimensions: 24cm x 18cm x 11cm (length x width x height)
Weight: 0.9 kg
Power supply: 4 Type AA alkaline batteries (not included)
Battery autonomy: 800hr.
Mains supply: 100-240 VAC, 50-60 Hz (Universal AC adapter included)
Inputs: BNC (pH or ORP electrode), pin-tip (reference electrode), 8-pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), shaker input (stirring probe)
Outputs: USB, RS232
Measuring modes: pH, mV, relative mV (RmV) or ORP with temperature; conductivity, TDS, salinity or resistivity with temperature
Calibration: pH with calibration edit option, relative mV (RmV), ORP, conductivity with calibration edit option, and temperature
Channels: 2
pH calibration points: 1 to 5 points
Conductivity calibration points: up to 5 points
Cell count: 0.001 to 199.9cm-1
Operating temperature: 5°C to 45°C (ambient)
Temperature probe calibration: 1-point temperature compensation calibration
Temperature selection: manual or automatic with ATC temperature probe
 
Parameter specifications:
pH:
Range: -2,000 to 20,000
Resolution: 0.1 / 0.01 / 0.001
Accuracy: ± 0.002
Calibration points: Up to 5 points.
Edit Calibration: Yes.
mV/RmV:
Range: ± 2000.0 mV
Resolution: 0.1
Accuracy: ± 0.2 mV or 0.05% of reading
REDOX mode: Yes
Conductivity:
Range: 0.001 µS to 3000 mS
Resolution: 0.001 µS
Accuracy: 0.5% of reading +1 digit
Temperature compensation: Linear
Compatible cell constants: 0.001 to 199.9
Calibration: 5 points
Edit Calibration: Yes
Resistivity:
Range: 2 ohm to 100 meg-ohm
Resolution 2 ohms-cm
Accuracy: 0.5% of reading +1 digit
Salinity:
Range: 0.01 to 80.0 ppt NaCl
Resolution: 0.01
Accuracy: +0.1
TDS:
Range: 1 to 200 ppt
Resolution: 4 significant digits
Accuracy: 0.5% of reading +1 digit
Temperature:
Range: -5 to 150 ºC (23 to 221 ºF )
Resolution: 0.1
Accuracy: ± 0.1
Offset calibration:1 point

pH / mV / RmV / Conductivity / TDS / Salinity / Temperature

pH electrode
Conductivity probe
Temperature probe
Calibration standards
Storage solution for pH electrode
pH buffers
Other accessories available on request