Moisture analyzer model ACCUPOINT 2

Description

Specifications

Parameters

Accessories

Descargas

The Accupoint 2 is an advanced moisture analyzer equipped with a microprocessor and an integrated two-stage pressure regulator. With a wide operating range from 50 to 3000 psig, it is perfectly suited for most industrial, natural and process gas flows, ensuring optimal and accurate performance. Characteristics:
  • Units of measure: microprocessor-based electronics allow a choice of display options, including ppmV, ppmW, lbs/mmscf or °C and °F dew point.
  • Three-button user interface: the Mode/Enter key, together with the simple Up and Down keys, make menu use quick and easy.
  • LCD display: the integrated digital display allows direct indication at the point of use and quick setting of control parameters in the field.
  • Scalable output: flexibility to change output scales in the field. No need to replace electronic components. Simply access the menu via the Mode/Enter key and select the output scale.
  • 2-wire or 3-wire modes: standard simplicity of a 2-wire loop-powered transmitter or additional RS-232 output function in 3-wire mode.
  • In-line verification: use the simple Delta Flow procedure to quickly verify the linearity and performance of the sensor in-line.
Available versions: * Accupoint LP2: Low pressure humidity version (5-100 psig)
Performance Accupoint 2
Operating range 0-1000 ppmV with 0.1 ppm resolution (100 cc flow units)

0-5000 ppmV with 1 ppm resolution (10 cc flow units)

Detection limit (LDL) 0.5 ppmV
Accuracy (greater than) Standard cells: 5% of reading or 0.4 ppm, whichever is greater.*
Operating temperature – 20 °C to +60 °C (-4 °F to 140 °F)
User interface 3-key touch panel
Display 1-line, 16-character, alphanumeric LCD, 3/8″ high digits
Display options ppmV, ppmW (requires user input of molecular weight), °C or °F dew point and lbs/mmscf

(Note: °C and °F dew point refer to atmospheric pressure.

Pressure dew point is available if the user enters the operating pressure).

Gas handling system and conditions
Gas connections 1/8″ compression
Inlet Pressure 50-3000 psig
Flow rate Sample: 10 sccm or 100 sccm

Bypass: 1000 sccm

Dimensions

Height x Width x Depth [inches (mm)].

7″ (17.8 cm) wide x 4½” (10.8 cm) deep x 8¼” (21.6 cm) high.
Weight 4.1 kg (9 lbs)
Electric
Alarm indicators User programmable 4-20 mA loop signal (2-wire mode)

Non-isolated 4-20 mA current sink (3-wire model) (no CSA approval)

Isolated RS232 in 3-wire mode only (optional, no CSA approval)

Power Requirements 24 VDC 20 % 2-wire loop powered (customer supplied)

24 VDC 10 % 3-wire common ground (customer supplied)

Humidity
Humidity in gases

Consult