Halogen and Sulfur Analyzer Model TOX-300

Description

Specifications

Parameters

Standards

Accessories

Descargas

Videos

Equipment for easy determination of total halogen/sulfur in process control, final product, environmental control and determination of the parameter AOX. For this purpose, a new software is available, which allows a simple and fast handling of the results. With the new SCP system for combustion, the combustion of any matrix can be carried out with a single combustion method. Characteristics:
  • Determination of total halogen/total sulfur by columbiometry.
  • Solid, liquid or gaseous samples.
  • Max. oven temperature 1100 ºC.
  • Electrodes: Silver electrode (chlorine), Platinum electrode (sulfur).
  • Detection range.
    • Chlorine: 0.05 - 50 µg.
    • Sulfur: 0.05 - 50 µg.
  • New SCP (Secure Combustion Program) system for all types of dies.
  • Connection to PC via USB port.
  • Deodorizer system (optional).
  • Possibility of performing AOX by double column method (option TX4-04 absorption module required).
Replaced by the new TOX 500 model
Models Chlorine and Sulfur Analyzer / Total Organic Halogen Analyzer model TOX-300
Method of analysis Oxidative decomposition / Columbimetric titration
Oxidative decomposition Quartz tube combustion method
Samples Liquid, solid and gaseous (using gas injector model GI-210)
Sample introduction Automatic sample introduction with a small vessel (ABC-210 unit)
Pyrolytic oven Horizontal oven: up to 1.100 ⁰C
Detection method Oxidation-reduction potential (potentiometric detection by potential)
Detector electrode Chlorine measurement: silver electrode Sulfur measurement: platinum electrode
Valuation control Automatic control of electrolytic current
Measurement ranges See Table of measuring ranges

 

Measurement ranges:

Absolute volume Solid samples Liquid samples Gaseous samples
Chlorine: 0.05 µg/ml to 50 µg 2 µg/g (30mg) 0.5 µg/ml (100 µl) 5mg/m3 (10ml)
Sulfur: 0.05 µg/ml to 50 µg 2 µg/g (30mg) 0.5 µg/ml (100 µl) 5mg/m3 (10ml)
Common samples: samples that are combustible below 1,100 ⁰C and have no interference.

 

Sample volume liquid samples: less than 100 µl, solid samples: less than 30 mg
Analysis time Less than 10 min/measurement (2µg measurements)
Temperature and humidity 15 to 35 ⁰C, below 80 % RH (Non-condensing)
Dimensions TOX-300 main unit: 550 x 360 x 437 mm (W x D x H)

ABC-210: 440 x 250 x 180 mm ( W x D x H )

Weight TOX-300 main unit: 36 Kg ABC-210: 11 Kg
Computer / O.S. Windows® 7 Preinstalled version English
Printer Windows® compatible printer
Gas Oxygen (purity greater than 99.7%) Argon (purity greater than 99.98 %)
Power supply TOX-300 main unit: AC 100 to 240 V, 50/60 Hz, 980 VA ABC-210: AC 100 to 240V, 50/60 Hz, 40 VA

AOF – PFAS

AOX – Adsorbable Organic Halogens

EOX – Extractable Organic Halogens

ISO 9562

ASTM D5808

ASTM D3120

ASTM D3246

ASTM D3961

EPA 9020B

EPA 9021

EPA 9076

ISO 11480

DIN 38414-17

DIN 38414- 18

Consult