INSTRU joins the United Nations Global Compact
Continue readingDescription
Specifications
Parameters
Accessories
The new Sievers Bioburden Rapid Analyzer model Soleil is based on a high-throughput flow cytometer to quantify bioburden in less than 45 minutes.
Compared to classical flow cytometry, we can analyze large fluid volumes(8 ml/min) almost in real time and, using a unique combination of viability staining, we can distinguish viable cells [bacteria and fungi (yeast and mold)] from non-viable ones (particles) that require days to be visible with classical methods.
The limit of detection is 1 cell/100 ml while the limit of quantification is <10 cells/100 ml of sample.
Simple sample matrices (e.g. WFI, PW, weak salt buffers) can be analyzed as is or with dilutions, thus solving any interference problems, while more complex ones (e.g. cell culture media) require evaluation based on their characteristic physicochemical properties with focused experiments.
The R&D team has used these applications in a variety of challenging matrices (e.g. enzymes, detergents, culture media, cosmetics).
The range of applications is wide as it can be used in different key processes to obtain a fast and accurate Bioburden result:
Soleil can be used both in the laboratory and at-line in production.
Operating specifications:
|
Range |
<10 biotics/100 ml - 10 000 biotics/100 ml |
|
Type of example |
Discrete random sample |
|
Screen reading |
Biotics/100 mL |
|
Calibration |
Normally stable for 12 months |
|
Analysis time |
< 45 minutes Total time to result (TTR) |
|
Instrument sample flow rate |
8 ml/min |
|
Sample temperature |
35-39 °C (95-102.2 °F) |
|
Ambient temperature |
17-30 °C (63-86 °F) |
|
Sample pressure |
N / A |
Analyzer specifications:
|
Exits |
USB port, Ethernet TCP/IP |
|
Display |
Touch screen for 14″ laptop (1920×1080) |
|
Feeding |
100-240 V~, 50-60 Hz, 165 VA |
|
Fuses |
2x T 5A 250 VAC Replace with fuse of the same type and size Device inlet 5×20 mm size |
|
Laser (Class 1) |
Wavelength: 488 nm Maximum power: 135 mW |
|
Dimensions |
Width: 48.3 cm (19 in.) Depth: 43.2 cm (17 inches) Height (without laptop): 43.2 cm (17 inches) |
|
Weight |
18.1 kg (40 lbs) |
|
Classification of the casing |
IP30 |
|
Safety certifications |
ETL, CE, UKCA, FCC, CSA 22.2 |
Environmental Specifications:
|
Maximum relative humidity |
0-85%, non-condensing |
|
Maximum altitude |
3,000 m (9,800 ft) |
|
Degree of contamination |
2 |
Bioburden – Bioburden
To be consulted
Pide información de Bioburden rapid bioburden analyzer “Bioburden”.
We want to play an active role in creating a safer and more environmentally respectful future.
Related resources
INSTRU joins the United Nations Global Compact
Continue reading
Dew Point measurement: how to control residual moisture in industrial processes
Continue reading
Our thanks for the Marta Mayordomo Award
Continue reading
New reference standards certified at –40 °C for Jet Fuel testing
Continue reading
InnovaPrep solutions for microbiological detection in critical environments
Continue reading
ASTM D93-25: Standard update to improve traceability of the flash point
Continue reading