ACR-M3mk2 carbonaceous micro carbonaceous residue tester for micro carbonaceous residue testing

Description

Specifications

Parameters

Standards

Accessories

  • The new mk2 model automates the vaporization / coking process of the carbonaceous micro residue test.
  • Built according to standards: ASTM D4530, ISO 10370
  • Automatic gas flow regulation.
Method of analysis:
Purge with N2 10 minutes
Thermal cycle with ramp up to 500°C (according to standard) in 30 minutes
Stabilized at 500ºC 15 minutes
Cooling
New model incorporating the latest generation microprocessor for heating ramps.
Programmable profiles. Ramps up to 15ºC/minute.
Digital display with indication of:
  • N2 flow rate
  • Temperature
  • Weather
Improved working times:
  • Conradson: Less than 1 hour
  • Ashes: Less than 3 hours (Protected at 600ºC by software)
Working range: ambient up to 500°C
Dimensions (WxDxH): 350 × 390 × 460 mm
Weight: 21 kg
Power supply: 100 / 120 V or 220 / 240V 50 / 60Hz

Micro Carbonaceous Residue

ISO 10370
ASTM D4530

Small, medium, large vials
Mixed vial holder
Other accessories on request

Pide información de ACR-M3mk2 carbonaceous micro carbonaceous residue tester for micro carbonaceous residue testing

We want to play an active role in creating a safer and more environmentally respectful future.